An MRI comparative image evaluation under diagnostic and radiotherapy planning set-ups using a carbon fibre tabletop for pelvic radiotherapy; Comparaison d'images d'IRM acquises à l'aide du positionnement pour le diagnostic ou la planification en radiothérapie pelvienne avec un support plat

Fecha de publicación: Fecha Ahead of Print:

Autores

  • Maria Del Rosario Pastor Juan

    Autor

  • Roberto Berenguer Serrano

    Autor

  • Elena Lozano Setién

    Autor

  • Ignacio Andres Garcia

    Autor

  • Maria Isabel Tercero Azorin

    Autor

  • Mar Sevillano Capellan

    Autor

Autores externos

  • E., Jimenez-Jimenez
  • Á., Rovirosa
  • E., Korte

Abstract

PURPOSE: To quantitatively and qualitatively evaluate and compare MR images acquired from the usual diagnostic set-up with images acquired in a radiotherapy set-up. MATERIALS AND METHODS: MR images of a phantom and 18 prostate cancer patients imaging using the usual diagnostic set-up with a curved tabletop (DX-set-up) and using a radiotherapy set-up (RT-set-up) with a flat tabletop, were analysed retrospectively. Quantitative assessments were made by measuring signal-to-noise ratio (SNR) and contrast to noise ratio (CNR). Non-parametric tests were used. Qualitative assessments were made independently by three radiologists. Inter-rater reliability was measured as the percentage of agreement and Gwet's AC2 test. RESULTS: The use of the RT-set-up was linked to 19.3% SNR drop on the phantom image. Up to 24% SNR prostate drop was observed in patients' images. Up to 34% CNR drop was observed in patients' images. Although image quality of the RT-set-up was acceptable, all measured domains were worse than the corresponding DX-set-up. Additionally diagnostic images were linked to a better percentage of agreement among raters. CONCLUSIONS: SNR, CNR and image quality were significantly worse when using the RT-set-up than with the DX-set-up although Image quality in the RT-set-up was acceptable.

Copyright © 2019 Société française de radiothérapie oncologique (SFRO). Published by Elsevier Masson SAS. All rights reserved.

Datos de la publicación

ISSN/ISSNe:
1278-3218, 1769-6658

Cancer Radiotherapie  ELSEVIER

Tipo:
Article
Páginas:
296-303
PubMed:
31133514
Enlace a otro recurso:
www.scopus.com

Citas Recibidas en Scopus: 2

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • CNR; Carbon fibre; Fibre de carbonne; IRM; MRI; Radiotherapy; Radiothérapie; SNR; Support plat; Tabletop

Compartir